Section 3: Using the System
Powering up the System
Note: These instructions assume that the appropriate personnel have been trained.
1. Ensure that the system battery is properly installed or that the system has been plugged into an appropriate power source.
2. Turn the system power ON by using the power switch on the rear of the unit.
3. The system will display an "Enter Password" prompt. Using the SCROLL control button, highlight each digit of the PEN number, pressing the ENTER button after each selection.
4. When finished entering the PIN number, SCROLL to highlight DONE, and press ENTER.
When shipped, the unit contains an administrator-level default password 1914.
Caution! This is an administrator-level password and must be changed immediately! Note that the system MUST be attached to a display and keyboard to accomplish this.
5. The system will display a "Date/Time" verification message. The user must verify the correct date and time and then press ENTER. If the date or time is incorrect, the user must follow the screen prompts to change them.
6. The system will then display a "Warming Up -Please Wait" message along with a status indicator of time remaining. The system's YELLOW LED will flash during this time, indicating that the user must wait.
7. Once warmed up, the system will automatically perform a calibration. During the calibration, the LED on the system will be YELLOW, indicating that the user must wait.
If the calibration fails, all of the LEDs will blink, indicating the calibration has failed. Please follow troubleshooting instructions for failed calibration (see Troubleshooting section).
During the warm up period, an Administrator can access the various menu functions and perform operations other than sampling.
When system is ready, it will display a "Ready" message and the GREEN LED will illuminate on the Analyzer Unit.